Title: Single-pass lead VDD pacing system in children and adolescents
Subtitle: Report of twenty-four patients
Author(s): ÇELIKER, Alpay , KARAKURT, Cemşit , KARAGÖZ, Tevfik
Journal: Acta Cardiologica
Volume: 61    Issue: 1   Date: February 2006   
Pages: 63-67
DOI: 10.2143/AC.61.1.2005141

Abstract :
The dual-chamber pacing systems allow for AV synchrony, but generally require the placement of two separate transvenous leads. Single-lead atrioventricular synchronous pacing system (VDD) using single-pass leads has been accepted as therapeutic alternative for atrioventricular block with normal sinus node function.
The aim of this study is to evaluate clinical performance of single-pass lead VDD pacing systems in
24 consecutive patients in a ten-year period.The study group includes 17 (70.8%) male and 7 (28.8%)
female patients. The mean age and mean weight during pacemaker implantation was 10.4±3.8 years (4-17 years) and 30±11kg (13.5-55kg), respectively. The patients have been followed for 7-84 months (median 42months). The percentage of atrial sensing-ventricular pacing changed from 75 to 99.9% at the last control.
During the follow-up period,pacing mode was changed to VVIR due to complete loss of AV synchrony in five patients (21%). Four of them had underlying cardiac disease. In these patients loss of AV synchrony might be related to cardiac enlargement/abnormal cardiac anatomy or small atrial dipole-ventricular tip length.
Despite the loss of AV synchrony in post surgical AV block or dilated cardiomyopathy, single-lead VDD pacing systems can be successfully used in children with complete AV block and normal sinus node function. Patients and lead selection should be taken into consideration for the maintenance of AV synchrony.